NKS-B GammaTest:Spectra for calibration
From Gamma Wiki
					
										
					
					The task is to report at least all identified nuclides - perfect solution includes the identification of all significant peaks, including background peaks i.e. no background subtraction. All the relevant spectra is here: Spectra for calibration
1. Peak/nuclide identification exercise spectrum file: 12711G6.chn
- Sample: sewage sludge sample in 0.5 l Marinelli beaker
 - Detector: coaxial, p-type HPGe detector, relative efficiency 39%
 - Energy calibration options
- spectrum: calibG6.chn, certificate of calibration source: certif.txt (files attached)
 - energy calibration equation: E = 0.3337 * ch – 0.183 (E= energy in keV, ch = channel)
 - channel-energy –pairs:
- 140.2 – 46.54 keV
 - 1531.4 – 511.0
 - 4378.6 – 1460.8
 - 7835.4 – 2614.5
 
 
 
2.	Peak/nuclide identification exercise; spectrum file: 11100B1.chn
- sample: contaminated film dosimeter
 - detector: Canberra BEGe detector BE5030, relative efficiency 50%, thin entrance window,thin Ge dead layer
 - energy calibration options
- energy calibration equation: E = 0.3305 * ch + 0.7301 (E= energy in keV, ch = channel)
 - channel – energy -pairs:
- 67.7 – 23.1 keV
 - 516.3 – 171.3
 - 740.2 – 245.4
 - 1258.3 – 416.6